FastCCD chiller =============== The CRYOMECH CP830 cryo-chiller is located near the FastCCD electronics rack. Below are intructions to start ip up and shut it down. .. |FCCD-chiller| image:: FastCCD-Cryomech-CP830-chiller.jpg :width: 100% .. |FCCD-water-line| image:: FastCCD-Cryomech-CP830-chiller-wall-water-line.jpg :width: 100% .. |FCCD-cryohead| image:: FastCCD-Cryomech-CP830-cryohead.jpg Start up: ~~~~~~~~~ +---+-------------------------+------------------------------------------------+ |1. | |FCCD-chiller| | make sure that the power is connected and the | | | | interlock allows the chiller operation (red LED| | | | on front panel under interlock connector is on)| +---+-------------------------+------------------------------------------------+ |2. | |FCCD-cryohead| | check that the two Helium lines are connected | | | | both to the chiller body and to the chiller | | | | head (on the beamline) | +---+-------------------------+------------------------------------------------+ |3. | |FCCD-cryohead| | check that the motor power cord is connected | | | | both to the chiller body and to the chiller | | | | head | +---+-------------------------+------------------------------------------------+ |4. | |FCCD-water-line| | Turn the cooling water outflow red valve (right| | | | of the green valve marked "FastCCD") | | | | anti-clockwise all the way | +---+-------------------------+------------------------------------------------+ |5. | |FCCD-water-line| | Open the cooling water inflow green valve | | | | marked "FastCCD" | +---+-------------------------+------------------------------------------------+ |6. | |FCCD-chiller| | close the circuit breaker on the chiller front | | | | panel | +---+-------------------------+------------------------------------------------+ |7. | |FCCD-chiller| | press the large green button on the front | | | | panel, the button will lighten up | +---+-------------------------+------------------------------------------------+ |8. | | the chiller should make a regular (~1Hz) loud | | | | 'chick' noise | +---+-------------------------+------------------------------------------------+ |9. | | the FastCCD temperature shoud start decreasing | | | | after ~1/2 hour | +---+-------------------------+------------------------------------------------+ Potential *DANGER* with wall water line: ---------------------------------------- Never close the cooling water outflow red valve while leaving the water cooling inflow green valve open, as this may lead to a water leak in the chiller. Shutdown: ~~~~~~~~~ +---+-------------------------+------------------------------------------------+ |1. | |FCCD-chiller| | press the large green button on the front | | | | panel. The chiller should fall silent and the | | | | green light go off | +---+-------------------------+------------------------------------------------+ |3. | |FCCD-chiller| | lower the circuit breaker on the chiller front | | | | panel | +---+-------------------------+------------------------------------------------+ |4. | |FCCD-water-line| | close the cooling water inflow green water | | | | valve marked "FastCCD" on the wall | +---+-------------------------+------------------------------------------------+ Based on `elog_1359`_. .. _elog_1359: https://in.xfel.eu/elog/SCS+Beamline/1359