- 1. Revisions
- 2. Introduction
- 3. Acronyms
- 4. The pnCCD Hardware
- 4.1. Overview
- 4.2. The 1 Megapixel pnCCD Detector at European XFEL
- 4.3. In-Vacuum Electronics
- 4.4. Front-End Electronics
- 4.5. MPOD Power Supply
- 4.6. Uninterrupted Power Supply
- 4.7. pnCCD PLC System
- 4.8. pnCCD Vacuum System
- 4.9. pnCCD Cooling System
- 4.10. pnCCD Interlock System
- 4.11. pnCCD Motion System
- 5. The Hardware Setup and Cable Connections
- 5.1. In-Vacuum Electronics Cable Connections
- 5.2. Cable Connections for the Front-End Electronics Outside Vacuum
- 5.3. Cable Connections for the MPOD System
- 5.4. The Vacuum System’s Hardware
- 5.5. Hardware Setup for the pnCCD Cooling System
- 5.6. pnCCD Interlock System Setup
- 5.7. pnCCD Motion System Setup
- 6. pnCCD Control Software
- 7. pnCCD Data Acquisition
- 8. pnCCD Operation; What to Do While on Shift
- 8.1. pnCCD Is at Room Temperature When You Start Your Shift
- 8.2. pnCCD Is Already Running When You Start Your Shift
- 8.3. Operating the pnCCD Front-End Electronics
- 8.4. Changing Trigger Times:
- 8.5. Saving the pnCCD Gain Setting
- 8.6. pnCCD Power
- 8.7. Operating pnCCD Chillers and Heaters
- 8.8. Processing Dark Runs
- 8.9. Calibrating/Recalibrating Acquired Data
- 8.10. Online Preview
- 8.11. Pixel Saturation Levels
- 8.12. Long Shutdown Periods
- 9. Troubleshooting
- 9.1. Maxwell Cluster Is Unaccessible
- 9.2. The pnCCD Gain Setting Is Wrong at the End of the Dark Report
- 9.3. Problems with the Calibration Webservice
- 9.4. Problems with Calibrating/Recalibrating Runs via Meta Data Catalog
- 9.5. Problems with the Online Preview
- 9.5.1. The Offset Time Stamp Is Not What It Should Be
- 9.5.2. The
Offset from
andNoise from
Fields on thepnCCD_Acquisition
Scene are Empty - 9.5.3. The Common Mode Corrected Image Is Stuck
- 9.5.4. The Offset Correction Karabo Device Is in Unknown State
- 9.5.5. Raw Image Is Stuck
- 9.5.6. Corrected Images are Stuck
- 9.6. Problems with the FastADCs
- 9.7. Errors While Powering Up/Down the pnCCD
- 9.8. Problems with the pnCCD Interlock System
- 9.9. A Bright Intense Stripe Is Observed in the Middle of pnCCD
- 9.10. When Is Annealing Necessary?
- 9.11. Problems with the pnCCD Motion System
- 9.12. Problems with the pnCCD Vacuum System
- 9.13. The Bottom pnCCD Module Is Not Cold Enough
- 9.14. Problems with the DAQ Syatem
- 9.15. The pnCCD Is Not Sensitive to the FEL
- 9.16. The Sequencer Shows Only One Running Counter Instead of Two
- 9.17. Problems with Karabo Bridge
- 10. Safety Measures