BeckhoffAgilentIonPump¶
Commands¶
| Key | Displayed Name | Description | Alias | Access Level | Allowed States |
|---|---|---|---|---|---|
| hvOff | hvOff | Turn the pump module OFF. This is done either
via serial command or digital output depending
on the firmware configuration. No channel
selection required as configured in the
firmware.
|
COff | EXPERT | |
| hvOn | hvOn | Turn the pump module ON. This is done either
via serial command or digital output depending
on the firmware configuration. No channel
selection required as configured in the
firmware.
|
COn | EXPERT |
Properties¶
| Key | Displayed Name | Description | Alias | Type | Access Level | Access Mode | Allowed States |
|---|---|---|---|---|---|---|---|
| controller.interlockStatus | Interlock status | Interlock status of the controller.
|
AInterlockStatus0 | String | OBSERVER | READONLY | |
| controller.setPointStatus | Setpoint status | Setpoint status of the controller.
|
ASetpointStatus0 | String | OBSERVER | READONLY | |
| currentCh1 | Current Ch.1 | Current for channel 1.
|
ACurrent1 | Float | OBSERVER | READONLY | |
| currentCh2 | Current Ch.2 | Current for channel 2.
|
ACurrent2 | Float | OBSERVER | READONLY | |
| currentCh3 | Current Ch.3 | Current for channel 3.
|
ACurrent3 | Float | OBSERVER | READONLY | |
| currentCh4 | Current Ch.4 | Current for channel 4.
|
ACurrent4 | Float | OBSERVER | READONLY | |
| currentSerial | Current | Current for the selected channel.
|
Float | OBSERVER | READONLY | ||
| isSwitchedOn | isSwitchedOn | Boolean indicating if the high voltage was
switched ON/OFF (valueBit 12) with the slot
calls hvOn and hvOff.
|
Bool | OBSERVER | READONLY | ||
| onSetpoint | On Setpoint | Boolean indicating if the pump has reached the
setpoint.
|
Bool | OBSERVER | READONLY | ||
| pressure | Analog pressure | Pressure read out from analog input. NOT
available in slim mode.
|
AAnalogPressure | Float | OBSERVER | READONLY | |
| pressureCh1 | Pressure Ch.1 | Pressure for channel 1. This is a serial
readout.
|
APressure1 | Float | OBSERVER | READONLY | |
| pressureCh2 | Pressure Ch.2 | Pressure for channel 2. This is a serial
readout.
|
APressure2 | Float | OBSERVER | READONLY | |
| pressureCh3 | Pressure Ch.3 | Pressure for channel 3. This is a serial
readout.
|
APressure3 | Float | OBSERVER | READONLY | |
| pressureCh4 | Pressure Ch.4 | Pressure for channel 4. This is a serial
readout.
|
APressure4 | Float | OBSERVER | READONLY | |
| pressureSerial | Pressure (serial) | Presssure of the selected channel; this is a
serial readout.
|
Float | OBSERVER | READONLY | ||
| slimInterface | Slim Interface | True: Slim interface with setpoint DI only.
False: Full interface as in IO struct.
|
Bool | OBSERVER | READONLY | ||
| voltage | Analog voltage | Voltage read out from analog input. NOT
available in slim mode.
|
AAnalogVoltage | Float | OBSERVER | READONLY | |
| voltageCh1 | Voltage Ch.1 | Voltage for channel 1.
|
AVoltage1 | Int16 | OBSERVER | READONLY | |
| voltageCh2 | Voltage Ch.2 | Voltage for channel 2.
|
AVoltage2 | Int16 | OBSERVER | READONLY | |
| voltageCh3 | Voltage Ch.3 | Voltage for channel 3.
|
AVoltage3 | Int16 | OBSERVER | READONLY | |
| voltageCh4 | Voltage Ch.4 | Voltage for channel 4.
|
AVoltage4 | Int16 | OBSERVER | READONLY | |
| voltageSerial | Voltage (serial) | Voltage for the selected channel; this is a
serial Property
|
Int16 | OBSERVER | READONLY | ||
| channelSelection | Channel Controller | Channel on the controller for the pump.
|
Int16 | EXPERT | READONLY | ||
| controller.baudRate | Baudrate | Baudrate: 1-4 => 1200-2400-4200-9600. This is a
controller only property.
|
ABaudrateHW0 | Int16 | EXPERT | RECONFIGURABLE | |
| controller.command | Serial Command | Send serial protocol command.
|
ASerial | String | EXPERT | RECONFIGURABLE | |
| controller.controlMode | Control Mode | Set the control mode. 0: Serial, 1: Remote, 2:
Local. This is a controller only property.
|
AControlMode0 | Int16 | EXPERT | RECONFIGURABLE | |
| controller.errorCode | Error code | Error code from the serial interface (hex).
|
AErrorcodeHW0 | Int16 | EXPERT | READONLY | |
| controller.errorString | Hardware error description | The actual translated error state of the
hardware. This is a serial property.
|
String | EXPERT | READONLY | ||
| controller.fixedStep | Fixed step | Fixed step setting of the controller.
|
AFixedStep0 | String | EXPERT | READONLY | |
| controller.hwStatus | Status | Hardware status of the controller.
|
AStatusHW0 | Int16 | EXPERT | READONLY | |
| controller.measurementUnit | Measurement unit | Measurement unit of the controller.
|
AMeasurementUnit0 | Int16 | EXPERT | RECONFIGURABLE | |
| controller.modelNo | Model no. | Model Number
|
AModelNo0 | String | EXPERT | READONLY | |
| controller.operatingMode | Operating Mode | Set the operating mode of the controller.
|
AOperatingMode0 | String | EXPERT | RECONFIGURABLE | |
| controller.operatingModeRead | Operating mode dummy | Operating mode: 0000000000 auto OFF; 0000000001
auto ON. This is serial read only property.
|
AOperatingMode0 | String | EXPERT | READONLY | |
| controller.protectMode | Protect mode | Protect mode of the controller.
|
AProtectMode0 | String | EXPERT | READONLY | |
| controller.serialNo | Serial No. | Serial number of the controller.
|
ASerialNo0 | String | EXPERT | READONLY | |
| controller.tempFan | Temperature at fan | Temperature measured at fan.
|
ATemperatureAtFan0 | Int16 | EXPERT | READONLY | |
| deviceInSlot | Device in Slot (serial) | Device in the selected channel; this is a
serial Property
|
Int16 | EXPERT | RECONFIGURABLE | ||
| deviceInSlot1 | Device in slot 1 | Device in slot 1.
|
ADeviceInSlot1 | Int16 | EXPERT | RECONFIGURABLE | |
| deviceInSlot2 | Device in slot 2 | Device in slot 2.
|
ADeviceInSlot2 | Int16 | EXPERT | RECONFIGURABLE | |
| deviceInSlot3 | Device in slot 3 | Device in slot 3.
|
ADeviceInSlot3 | Int16 | EXPERT | RECONFIGURABLE | |
| deviceInSlot4 | Device in slot 4 | Device in slot 4.
|
ADeviceInSlot4 | Int16 | EXPERT | RECONFIGURABLE | |
| highVoltageOn | HighVoltage On | Turn on high voltage of the selected channel;
this is a serial Property
|
Bool | EXPERT | RECONFIGURABLE | ||
| highVoltageOn1 | HighVoltage 1 On | Turn on high voltage of channel 1.
|
AHighVoltageOn1 | Bool | EXPERT | RECONFIGURABLE | |
| highVoltageOn2 | HighVoltage 2 On | Turn on high voltage of channel 2.
|
AHighVoltageOn2 | Bool | EXPERT | RECONFIGURABLE | |
| highVoltageOn3 | HighVoltage 3 On | Turn on high voltage of channel 3.
|
AHighVoltageOn3 | Bool | EXPERT | RECONFIGURABLE | |
| highVoltageOn4 | HighVoltage 4 On | Turn on high voltage of channel 4.
|
AHighVoltageOn4 | Bool | EXPERT | RECONFIGURABLE | |
| maximumPower | Max. Power | Maximum power for the selected channel; this is
a serial Property
|
Int16 | EXPERT | RECONFIGURABLE | ||
| maximumPower1 | Max. Power Ch.1 | Maximum power for channel 1
|
AMaximumPower1 | Int16 | EXPERT | RECONFIGURABLE | |
| maximumPower2 | Max. Power Ch.2 | Maximum power for Channel 2
|
AMaximumPower2 | Int16 | EXPERT | RECONFIGURABLE | |
| maximumPower3 | Max. Power Ch.3 | Maximum power for Channel 3
|
AMaximumPower3 | Int16 | EXPERT | RECONFIGURABLE | |
| maximumPower4 | Max. Power Ch.4 | Maximum power for Channel 4
|
AMaximumPower4 | Int16 | EXPERT | RECONFIGURABLE | |
| protectCurrent | Protect current | Protect current for the selected channel; this
is a serial Property
|
Int16 | EXPERT | RECONFIGURABLE | ||
| protectCurrent1 | Protect current Ch.1 | Protect current for channel 1.
|
AProtectCurrent1 | Int16 | EXPERT | RECONFIGURABLE | |
| protectCurrent2 | Protect current Ch.2 | Protect current for channel 2.
|
AProtectCurrent2 | Int16 | EXPERT | RECONFIGURABLE | |
| protectCurrent3 | Protect current Ch.3 | Protect current for channel 3.
|
AProtectCurrent3 | Int16 | EXPERT | RECONFIGURABLE | |
| protectCurrent4 | Protect current Ch.4 | Protect current for channel 4.
|
AProtectCurrent4 | Int16 | EXPERT | RECONFIGURABLE | |
| setPoint | Setpoint | Setpoint for the selected channel; this is a
serial Property
|
Float | EXPERT | RECONFIGURABLE | ||
| setPoint1 | Setpoint Ch.1 | Setpoint for channel 1.
|
ASetpoint1 | Float | EXPERT | RECONFIGURABLE | |
| setPoint2 | Setpoint Ch.2 | Setpoint for channel 2.
|
ASetpoint2 | Float | EXPERT | RECONFIGURABLE | |
| setPoint3 | Setpoint Ch.3 | Set Point for channel 3.
|
ASetpoint3 | Float | EXPERT | RECONFIGURABLE | |
| setPoint4 | Setpoint Ch.4 | Setpoint for channel 4.
|
ASetpoint4 | Float | EXPERT | RECONFIGURABLE | |
| targetVoltage | Target Voltage | Target Voltage for the selected channel; this
is a serial Property
|
Int16 | EXPERT | RECONFIGURABLE | ||
| targetVoltage1 | Target Voltage Ch.1 | Target Voltage Channel 1
|
ATargetVoltage1 | Int16 | EXPERT | RECONFIGURABLE | |
| targetVoltage2 | Target Voltage Ch.2 | Target voltage for channel 2
|
ATargetVoltage2 | Int16 | EXPERT | RECONFIGURABLE | |
| targetVoltage3 | Target Voltage Ch.3 | Target voltage for channel 3
|
ATargetVoltage3 | Int16 | EXPERT | RECONFIGURABLE | |
| targetVoltage4 | Target Voltage Ch.4 | Target voltage for channel 4.
|
ATargetVoltage4 | Int16 | EXPERT | RECONFIGURABLE | |
| temperatureHV | Temperature HV | Temperature for HV end for the selected
channel; this is a serial Property
|
Int16 | EXPERT | READONLY | ||
| temperatureHV1 | Temperature HV Ch.1 | Temperature HV for channel 1.
|
ATemperatureHV1 | Int16 | EXPERT | READONLY | |
| temperatureHV2 | Temperature HV Ch.2 | Temperature HV for channel 2.
|
ATemperatureHV2 | Int16 | EXPERT | READONLY | |
| temperatureHV3 | Temperature HV Ch.3 | Temperature HV for channel 3.
|
ATemperatureHV3 | Int16 | EXPERT | READONLY | |
| temperatureHV4 | Temperature HV Ch.4 | Temperature HV for channel 4.
|
ATemperatureHV4 | Int16 | EXPERT | READONLY |