End of Shift Procedure¶
This section describes the procedure to put the instrument in safe conditions at the end of a shift.
- SA3 beamline
- Stop mono scan.
- Stop mono gap sync .
- Set number of pulses to 1.
- Stop injection to GATT (if the next injection is not expected to happen in the next 1 hour or more).
- Insert pre-absorber.
- Extract TPI.
- PP laser
- Close the laser shutter.
- Turn the key to disable the laser interlock.
- Set the number of pulses to 1.
- Stop delay scan.
- Turn OFF the LT3 motor.
- Stop the online monitoring tool for fine timing.
- SCS beamline
- Close all beam line gate valves while monitoring the pressure in each section.
- Insert B4C at DPS-2.
- MCPs or APDs (TIM, CHEM, XOX…)
- Set all MCP and APD HV to zero and turn them OFF.
- XGM / HAMP
- Set voltages of the HAMP h HV MP to zero, power off the corresponding channel.
- DSSC
- Turn off the DSSC following the appropriate procedure.
- CHEM
- Open filter towards hRIXS
- Close valve towards hRIXS
- XRD
- Open filter towards hRIXS
- Close valve towards hRIXS
- Turn off all XRD motors.
- hRIXS
- Close all valves between vaccuum sections and towards bellows
- Stop camera
- DAQ
- Ignore data (in Run Controller scene) both in SCS and SA3 topics.
- Radiation interlock
- Make sure the shutter can be safely closed.
- Close the shutter.
- Break the radiation interlock.