TIM
: The Transmission Intensity Monitor¶
The transmission intensity monitor (TIM) is used to detect the intensity after the sample. It is used either for normalization or to measure x-ray absorption scans.
The TIM detects primarily the x-ray fluorescence coming out a CVD
diamond film or a YAG crystal that is used as active instrument beam
stop. For low intensities, the fluorescence detection could change
into photoelectron detection. Furthermore, for particular experiments
with very intense beams when the IBS would be damaged by each shot
resulting in a non-linear fluorescence intensity measurement, a gas
will be used as source of fluorescence or photoelectron. The measuring
principle is displayed in tim_scheme
below. The x-ray beam
with sizes varying between the blue and green patterns impinges in the
diamond film at 45° and ejects both electrons and fluorescence photons
that can be detected with the MCP.
The detection system is a MCP and its charge is integrated, amplified, digitized and measured with a micro-TCA. The possibility of detecting both fluorescence and electrons can be realized by means of the potentials to apply to the MCP and anode. Alternativly a grid can be put in front of the MCP to repel charged particles in the fluorescence detection mode. This allows using always the same MCP electric configuration.
Several MCPs can be installed in the chamber, as well as imagers to
detect the beam on the diamond or YAG screen. The MCPs and imagers
look at the screen at different angles, as shown in
tim_chamber
.
The CVD and the YAG are mounted on a translation stage that allows to move them in and out and change from one to the other. At the exit port of the chamber, a B4C beam stop is installed.
Interlocks¶
The roughing vacuum is shared with the FCCD, so we better do not open the roughing vacuum while they are pumping down.
No | Property | Op | Value |
---|---|---|---|
c1 | SCS_CDIDET_FCCD2M/GAUGE/P9.Value | GT | 1e-2 |
c2 | SCS_CDIDET_FCCD2M/VALVE/SRV2.state.IsClosed | EQ | FALSE |
c3 | SCS_BLD_TIM/TPUMP/TP1.state.RPM80 | EQ | FALSE |
No | Logic | Action |
---|---|---|
a1 | c1 or (c2 and c3) | CClose |
No | Logic | Action |
---|---|---|
a1 | c1 or (c2 and c3) | COff |
Like always, if the turbo pumps fail, their gate valve should close. Unless, well, the vacuum is really bad and the pump is not running then we probably want to pump down, as described for the DPS.
No | Property | Op | Value | Hyst | Timeout |
---|---|---|---|---|---|
c1 | SCS_BLD_TIM/TPUMP/TP1.state.RPM80 | EQ | FALSE | ||
c2 | SCS_BLD_TIM/GAUGE/PCC22 | LT | 1e-4 | 1 | 100000 |
No | Logic | Action |
---|---|---|
a1 | c1 and c2 | CClose |