Electromagnet before 2019-09-10¶
This page lists the parameters of the first version of the FFT electromagnet that was installed in the FFT end-station from the start of SCS operation up to the 10th of September 2019 when an improved version was put in place instead.
Voltage | Current | Field |
---|---|---|
V | A | mT |
0 | 0 | 4 |
1 | 1.3 | 50 |
2 | 2.6 | 98 |
3 | 4 | 150 |
4 | 5.3 | 200 |
5 | 6.5 | 250 |
6 | 7.7 | 283 |
7 | 9 | 315 |
8 | 10 | 335 |
9 | 11.1 | 355 |
10 | 12.2 | 362 |
The magnetic field generated by the electromagnet is shown in
electromagnet_field
.
gap between pole (mm) | maximum current (A) | maximum field (T) |
---|---|---|
7.7 | 12 | 0.37 |
The temperature reached near the coil winding edge on the pole is shown in
electromagnet_temperature
. Combination of 5°C cooling and high water
flow-rate should keep the coil temperature <50°C when operating at 12 A in
vacuum.
The geometry of the pole shoe is defined in CAD file.